OOISensors
Fiber Optic Sensors System
Installation and Operation Manual
Document Number FOXY-AL300-000-02-0207
Offices: Ocean Optics, Inc. World Headquarters
830 Douglas Ave., Dunedin, FL, USA 34698
Phone 727.733.2447
Fax 727.733.3962
8 a.m.– 8 p.m. (Mon-Thu), 8 a.m.– 6 p.m. (Fri) EST
E-mail: [email protected] (General sales inquiries)
[email protected] (Questions about orders)
[email protected] (Technical support)
Zobrazit stránku 0
1 2 3 4 5 6 ... 137 138

Shrnutí obsahu

Strany 1 - Fiber Optic Sensors System

► OOISensors Fiber Optic Sensors System Installation and Operation Manual Document Number FOXY-AL300-000-02-0207 Offices: Ocean Opti

Strany 2

Table of Contents Product-Related Documentation • MultiFrequency Phase Fluorometer Installation and Operation Manuall • Various USB spectrometer d

Strany 3 - Table of Contents

6: Hardware Data Sheets and Instructions Oxygen Probe Temperature Control Accessories Temperature affects the fluorescence decay time, the fluoresce

Strany 4

6: Hardware Data Sheets and Instructions 6. Locate the port on the module that is connected to the thermistor in use. Note The RS-232 module can s

Strany 5 - FOXY-AL300-000-02-0207 iii

6: Hardware Data Sheets and Instructions Operation ► Procedure 1. Connect the FOXY-TK1 to the 10-foot wire (model number FOXY-T-WIRE) that comes w

Strany 6

6: Hardware Data Sheets and Instructions Specification Value Data acquisition FOXY-T-MOD-K RS-232 module Probe material Chomega-Alomega 304 stainl

Strany 7 - FOXY-AL300-000-02-0207 v

6: Hardware Data Sheets and Instructions Specification Value Element type Platinum 100 ohm Single 3 Wire Sheath material 316 SS Light Sources The

Strany 8

6: Hardware Data Sheets and Instructions Using the Continuous Mode ► Procedure 1. Turn the switch on the back of the LS-450 to Contin. This means

Strany 9 - About This Manual

6: Hardware Data Sheets and Instructions • When using the pulsed flash mode, you need to ensure that a constant number of flashes occur for every

Strany 10 - Upgrades

6: Hardware Data Sheets and Instructions R-LS-450 Rack Mount Blue LED Pulsed Light Source The R-LS-450 is shipped with the following jumper pin co

Strany 11 - Sensors System Overview

6: Hardware Data Sheets and Instructions Note This feature is only available with an ADC1000 A/D converter and with a J-series or later version of t

Strany 12 - HIOXY Probes

6: Hardware Data Sheets and Instructions R-LS-450 Operating Matrix The following table provides information on configuring the jumper blocks on the

Strany 13 - Software

Chapter 1 Sensors System Overview Overview The following sections provide a general overview of the Oxygen Sensors System and the pH Sensors System.

Strany 14 - Documentation

6: Hardware Data Sheets and Instructions • The switch is turned to pulsed mode • There is a jumper over the pins in JP1 of the R-LS-450 • There is a

Strany 15 - OOISensors Software

6: Hardware Data Sheets and Instructions Pins on the JP3 (of the S2000) Integration time for DAQ700 (multiple of) Integration time for SAD500 (multi

Strany 16 - Installing OOISensors

6: Hardware Data Sheets and Instructions USB-LS-450 Temperature Measurement Specifications Specification Value Precision Better than 0.1° C Accurac

Strany 17 - FOXY-AL300-000-02-0207 7

6: Hardware Data Sheets and Instructions Use Notes • Gently remove the plastic cover from the SMA connector before use. Pulling the SMA connector aw

Strany 18 - Configure Menu Functions

6: Hardware Data Sheets and Instructions FOXY/HOXY/FOSPOR General Specifications Specification FOXY HIOXY FOSPOR Principle Photoluminescence-quenc

Strany 19 - Configure Hardware

6: Hardware Data Sheets and Instructions FOSPOR Specifications Parameter Dissolved Oxygen in Liquid Oxygen Gas (at 1 atmosphere) Sensor mechanism

Strany 20

6: Hardware Data Sheets and Instructions FOXY/MFPF Specifications Specification Dissolved Oxygen in H2O Oxygen Gas (at 1 atmosphere) Sensor mechan

Strany 21 - FOXY-AL300-000-02-0207 11

6: Hardware Data Sheets and Instructions HIOXY/MFPF Specifications Specification Dissolved Oxygen in Liquid Oxygen Gas (at 1 atmosphere) Sensor me

Strany 22 - FOXY (oxygen)

6: Hardware Data Sheets and Instructions 108 FOXY-AL300-000-02-0207

Strany 23 - Option Name Description

Chapter 7 Troubleshooting Overview The following sections contain information that may assist you in troubleshooting some of the more common problems

Strany 24

1: Sensors System Overview FOXY Probes FOXY oxygen sensors are designed for monitoring oxygen partial pressure in gas and aqueous solutions. Stand

Strany 25 - FOXY-AL300-000-02-0207 15

7: Troubleshooting Issue Possible Cause Resolution Peaks displayed in the graph screen are in incorrect places. Wavelength coefficients may be incor

Strany 26 - 16 FOXY-AL300-000-02-0207

7: Troubleshooting Issue Possible Cause Resolution High response times when measuring Samples Sample condition may be adversely affecting response

Strany 27 - Pressure

7: Troubleshooting Remove the Unknown Device from Windows Device Manager ► Procedure 1. Open Windows Device Manager. Consult the Windows operating

Strany 28 - Timing Tab

7: Troubleshooting 2. Delete the OOI_USB.INF in the INF directory. If your computer is running either the Windows 2000 or XP operating system, you m

Strany 29

7: Troubleshooting 114 FOXY-AL300-000-02-0207

Strany 30 - Display Tab

Appendix A External Triggering Ocean Optics S2000-series spectrometers, when used with OOISensors Software, provide two methods of acquiring data: no

Strany 31 - Log Tab

A: External Triggering 116 FOXY-AL300-000-02-0207

Strany 32 - Display Functions

Appendix B Chemical Effects on Oxygen Probes Summary of Chemical Effects on Oxygen Probes (Data Available upon Request) Chemicals Known to be Harmfu

Strany 33 - Subtract Dark

B: Chemical Effects on Oxygen Probes Chemical Supporting Data Available Gasoline (even in headspace) Yes N-Vinyl-2-Pyrrolidinone (even at 2.5%) Ye

Strany 34 - Data Acquisition Parameters

Appendix C pH Sensor Theory of Operation and Calculations Overview The following sections contain the equations that express the theory of pH sensor

Strany 35 - Cursor Functions

1: Sensors System Overview Hardware The following hardware is included with the Fiber Optic Sensors System or is compatible with the Fiber Optic Sen

Strany 36 - Cursor Bar

C: pH Sensor Theory of Operation and Calculations Model # 1 Looks at base peak only. No baseline correction. Uses nominal pK and slope. HighpHAb

Strany 37 - Spectral Graph

C: ph Sensor Theory of Operation and Calculations correct for baseline aselineAbsLowBufferBAAbsLowBufferAbsLowBufferA−=−−][][ sBaselineAbHighBuffer

Strany 38 - Temperature Chart

C: pH Sensor Theory of Operation and Calculations Model # 4 In addition to previous, it uses both acid and base peaks LowBufferpHLowH−=10 HighBuffer

Strany 39 - Time Chart Switch

C: ph Sensor Theory of Operation and Calculations Revised Model #4 Looks at base peak only. Corrects for baseline. lineAbsHighpHBaseAsBaselineAb

Strany 40 - File Menu Functions

C: pH Sensor Theory of Operation and Calculations 124 FOXY-AL300-000-02-0207

Strany 41 - FOXY-AL300-000-02-0207 31

Index A Accessories, 66 Bifurcated Optical Fiber Assemblies, 71 Overcoats, 68 Probes, 66 Sol-Gel Coated Membranes, 69 Temperature Compensation, 69 Au

Strany 42 - Spectrometer Menu Functions

Index F Factory Calibration, 48 Factory Calibration Services, 88 File Menu, 31 Exit, 32 Open Spectrum, 32 Open Time Chart, 32 Page Setup, 32 Print S

Strany 43 - Pulse LED

Index P Packing List, 4 Page Setup, 32 pH, 1 Films, 58 pH Probes, 58 RFP200-UV-VIS Reflective Film Probe, 59 TP300-UV-VIS Probe, 58 pH Sensors, 58 C

Strany 44 - Calibrate Menu Functions

Index 128 FOXY-AL300-000-02-0207

Strany 45 - FOXY-AL300-000-02-0207 35

1: Sensors System Overview Documentation The Fiber Optic Sensors System contains various documentation components on paper, floppy diskette and CD

Strany 46 - 36 FOXY-AL300-000-02-0207

Chapter 2 OOISensors Software OOISensors Overview OOISensors Software is Ocean Optics’ next generation of operating software for the Fiber Optic Sens

Strany 47 - Oxygen Sensors

2: OOISensors Software Installing OOISensors Before installing OOISensors Software, make sure that no other applications are running. ► Procedu

Strany 48 - Fluorescence Quenching

2: OOISensors Software 3. Click Next at the Welcome screen. The Choose Destination Location screen appears. 4. Accept the default or select another

Strany 49 - FOXY-AL300-000-02-0207 39

2: OOISensors Software Configuring OOISensors Software Once OOISensors is installed and the system has been rebooted, connect the spectrometer to t

Strany 50 - 40 FOXY-AL300-000-02-0207

2: OOISensors Software Configure Hardware The Configure | Hardware dialog box allows you to specify the hardware parameters for the spectrometer. Ty

Strany 51 - Theory - Henry’s Law

Additional Offices: Ocean Optics Asia 137 Xianxia Road, Suite 1802, Changning District, Shanghai, PRC. 200051Phone 86.21.5206.8686 Fax 86.21.

Strany 52 - Temperature

2: OOISensors Software If you choose the ADC1000-USB A/D Converter: Specify the serial number of the ADC1000-USB, or select the First Available ADC

Strany 53 - Samples to Use

2: OOISensors Software The System Type field in the upper, right corner allows you to select either Spectrometer or MFPF for the attached hardware.

Strany 54 - Calibrating Oxygen Sensors

2: OOISensors Software Sensors Tab for Spectrometer Systems Click on the Sensors tab in the Configure System screen to access Sensor controls. The

Strany 55 - Calibrating Standards

2: OOISensors Software Option Name Description Choose a spectrometer channel to view its sensor parameters Selects the spectrometer channel for wh

Strany 56 - Finalizing Calibration

2: OOISensors Software Option Name Description Pressure Compensation CH Monitors and corrects for pressure fluctuations in the sample (if you have

Strany 57 - About Factory Calibration

2: OOISensors Software Option Name Description Temperature Measurement Specifies the method used to monitor and correct for temperature fluctuation

Strany 58 - ► Procedure

2: OOISensors Software pH Use this screen to set parameters for a pH sensor. 16 FOXY-AL300-000-02-0207

Strany 59 - FOXY-AL300-000-02-0207 49

2: OOISensors Software Pressure Use this screen to set parameters for a pressure sensor. FOXY-AL300-000-02-0207 17

Strany 60

2: OOISensors Software CO2 Use this screen to enter parameters for a CO2 sensor. Timing Tab Click on the Timing tab in the Configure System screen

Strany 61

2: OOISensors Software The Timing tab contains the following options: Option Name Description Preset Duration Specifies the length of the entire t

Strany 62 - 52 FOXY-AL300-000-02-0207

Table of Contents About This Manual ... 7 Doc

Strany 63

2: OOISensors Software Option Name Description External Software Trigger: This mode allows data acquisition to be synchronized with external event

Strany 64

2: OOISensors Software Option Name Description Spectral Graph & Time Chart: The Spectral Graph appears in the top of the display window and the

Strany 65 - Calibration Data

2: OOISensors Software The Log tab contains the following options: Option Name Description Store to disk every X acquisitions Specifies the numbe

Strany 66 - Recalibration

2: OOISensors Software ► Procedure 1. Select the Single or Cont. button from the Scan section of the OOISensors screen in the lower right hand corn

Strany 67 - Chapter 4

2: OOISensors Software OOISensors always subtracts the dark measurement from the sample measurement in the OOISensors algorithm. Data Acquisition P

Strany 68 - Start Up

2: OOISensors Software Average This parameter specifies the number of spectra to average for the specified spectrometer channel. A higher value for

Strany 69 - FOXY-AL300-000-02-0207 59

2: OOISensors Software Magnification Options There are several magnification options from which to choose. The function chosen remains in use until

Strany 70 - 60 FOXY-AL300-000-02-0207

2: OOISensors Software X and Y Values The cursor taskbar displays the X value and Y value of the cursor point. Cursor Properties To the right of t

Strany 71 - Calibration

2: OOISensors Software Temperature Chart The Temperature Chart, when enabled, displays in the upper half of the OOISensors main screen. When this o

Strany 72 - 62 FOXY-AL300-000-02-0207

2: OOISensors Software ► Procedure To save the Temperature Chart without displaying the chart: 1. Select Configure from the OOISensors menu bar. 2

Strany 73 - FOXY-AL300-000-02-0207 63

Table of Contents Sensors Tab for Spectrometer Systems ... 11 pH ...

Strany 74 - 64 FOXY-AL300-000-02-0207

2: OOISensors Software Log On/Off Switch You can enable or disable the saving of data to a log file with the Log switch. Click on the switch to ena

Strany 75 - Accessories

2: OOISensors Software Save Time Chart This menu option saves the current time chart as a tab-delimited ASCII file. You can then open these files as

Strany 76 - Part Number Description

2: OOISensors Software Clear Spectrum Graph This option removes static spectra overlays from the graph. Clear Time Chart This option clears the ti

Strany 77 - Overcoats

2: OOISensors Software Scan This option takes a scan of your sample. When in Single mode, this function acts as a snapshot. When the button is dep

Strany 78 - Coverslips, and Slides

2: OOISensors Software Calibrate Menu Functions The following section provides information on options available from the Spectrometer menu. Optio

Strany 79 - Spectrometers

2: OOISensors Software Notes Be sure to retain the Wavelength Calibration Data. You will need this information if you reinstall OOISensors. If you

Strany 80 - Bushing

2: OOISensors Software 36 FOXY-AL300-000-02-0207

Strany 81 - Instructions

Chapter 3 Oxygen Sensors The following section provides information on operation of the oxygen sensors, as well as the theory behind sensor operati

Strany 82 - Measurement Range

3: Oxygen Sensors Fluorescence Quenching The following paragraph explains the principle of fluorescence quenching as it relates to oxygen sensors. O

Strany 83 - Chemical Compatibility

3: Oxygen Sensors Calibration Requirements for Oxygen Sensors The following section explains the principles of calibration for oxygen sensors, as we

Strany 84 - Cleaning Oxygen Probes

Table of Contents Chapter 3: Oxygen Sensors ...37 How the Oxygen Sensors Work ...

Strany 85 - Oxygen Probe Specifications

3: Oxygen Sensors The Stern-Volmer constant (k) is primarily dependent on the chemical composition of the ruthenium complex. Ocean Optics probes hav

Strany 86 - 76 FOXY-AL300-000-02-0207

3: Oxygen Sensors Compensating for Temperature with the Second Order Polynomial Algorithm If you decide to compensate for temperature, the relation

Strany 87 - FOXY-R Specifications

3: Oxygen Sensors T (ºC) T* (T/100K) Mole Fraction of oxygen in water at 1 atmosphere pO2 Weight Fraction (ppm) at 1 atmosphere pO2 (pure O2)

Strany 88 - FOXY-AL300 Specifications

3: Oxygen Sensors Samples should be maintained at a constant (+/-3° C*) temperature. If this is not practical, calibrate the Fiber Optic Sensors Sys

Strany 89 - FOXY-AL300-000-02-0207 79

3: Oxygen Sensors Calibrating Oxygen Sensors The following sections provide instructions on calibrating your oxygen sensors both with and without te

Strany 90 - Optic Probes

3: Oxygen Sensors 9. Select the algorithm type from the Curve Fitting drop-down menu. This algorithm will be used to calibrate your sensor system.

Strany 91 - FOXY-18G/21G Specifications

3: Oxygen Sensors 5. Once you click the green Scan Standard button, a red Scanning button appears. Watch the values in the Intensity column. When t

Strany 92

3: Oxygen Sensors Calibrating Oxygen Sensors with Temperature Compensation Perform the steps below to calibrate your oxygen sensors using temperatur

Strany 93 - FOXY-AL300-000-02-0207 83

3: Oxygen Sensors 3. Adjust the integration time so that the fluorescence peak does not exceed 3500 counts. Set the integration time to powers of t

Strany 94 - Probe with Light Shield

3: Oxygen Sensors 6. Get the proper temperature reading. Either: • Select Spectrometer | Sample Temperature from the menu bar in the Single Tempera

Strany 95 - Specification Value

Table of Contents Sol-Gel Coated Membranes, Microscope Coverslips, and Slides ... 68 Spectrometers ...

Strany 96 - FOXY-RESP Respiration Probe

3: Oxygen Sensors Note If the probe has an overcoat, then the standards used in the calibration can be of a state of matter different from the sampl

Strany 97 - Factory Calibration Services

3: Oxygen Sensors 2. Set the integration time for the entire calibration procedure when the probe is measuring the standard with zero concentration

Strany 98 - Cleaning

3: Oxygen Sensors 4. Place your cursor under the oxygen concentration of the first standard. 5. Click the green Scan Standard button or select S

Strany 99 - Overcoat Specifications

3: Oxygen Sensors Finalizing Calibration without Using Factory Calibration – Part 1 ► Procedure 1. Click the green Curve Fit button. A graph displ

Strany 100 - FOXY-TS1 Omega Thermistor

3: Oxygen Sensors Note In order to use the Second Order Polynomial algorithm, you must have had at least three standards of known oxygen concentrati

Strany 101 - Application Tips

3: Oxygen Sensors 6. Click the green Scan Standard button or select Spectrometer | Scan Standard from the menu bar. Note Enabling the optional Cont

Strany 102 - FOXY-TK1 Specifications

3: Oxygen Sensors • chXFoxy.cal The OOISensors.cfg file is the application configuration file. Calibration data is called from this binary file each

Strany 103 - FOXY-TK1-W Specifications

Chapter 4 pH Sensors Overview The Fiber Optic pH Sensor system consists of a fiber optic probe designed to hold immobilized indicator dye materials

Strany 104 - Light Sources

4: pH Sensors Note You can also use an RT-2-10 or RT-10-20 transflection tip, which enables the TP300-UV-VIS to be used for routine transmission m

Strany 105 - Using the Pulsed Mode

4: pH Sensors 2. Activate channels with pH sensors and enter their analysis wavelengths (table on next page): • HA (acid peak) • A (base peak) • B

Strany 106 - LS-450 Specifications

Table of Contents FOXY-RESP Specifications ... 86 Factory

Strany 107 - Jumper Block 1 (JP1)

4: pH Sensors Note Be careful not to over-tighten the tip. Over-tightening the tip can break the mirror in the transmission probe or break the film

Strany 108 - Jumper Block 3 (JP3)

4: pH Sensors Calibration The following sections contain information on calibrating your Fiber Optic pH Sensor system. Calibration requires recordin

Strany 109 - Using JP3 on the S2000

4: pH Sensors • Integration Period, Average, and Boxcar Smoothing are displayed for the active channel. These values can be modified and will be us

Strany 110 - Setting the Integration Time

4: pH Sensors 9. Click on the Update Channel Calibration button. • From this point on, when you click the Scan button, OOISensors displays intens

Strany 111 - USB-LS-450

4: pH Sensors 64 FOXY-AL300-000-02-0207

Strany 112 - 102 FOXY-AL300-000-02-0207

Chapter 5 Compatible Products and Accessories Overview The following sections provide ordering information for products and accessories compatible

Strany 113 - Use Notes

5: Compatible Products and Accessories Part Number Description FOXY-OR125-GT 1000-µm core diameter titanium fiber optic probe, 1/8" outer diam

Strany 114 - 104 FOXY-AL300-000-02-0207

5: Compatible Products and Accessories Part Number Description HIOXY-AL300 300-µm fiber probe in aluminum jacketing, 1m HIOXY-OR125 1000-µm fib

Strany 115 - FOSPOR Specifications

5: Compatible Products and Accessories Part Number Description FOXY-AF RTV healthcare-grade silicone overcoat for oxygen probes FOXY-AF-MG RTV hi

Strany 116 - FOXY/MFPF Specifications

5: Compatible Products and Accessories Part Number Description FOXY-GF Sol-gel coated glass fiber membranes (pack of 5) for testing viability of

Strany 117 - HIOXY/MFPF Specifications

Table of Contents Appendix B: Chemical Effects on Oxygen Probes ...117 Summary of Chemical Effects on Oxygen Probes ...

Strany 118 - 108 FOXY-AL300-000-02-0207

5: Compatible Products and Accessories Bifurcated Optical Fiber Assemblies and Splice Bushing http://www.oceanoptics.com/products/premgradebif.aspOc

Strany 119 - Troubleshooting

Chapter 6 Hardware Datasheets and Instructions The following section contains product data and operating instructions for hardware that is specific

Strany 120 - 110 FOXY-AL300-000-02-0207

6: Hardware Data Sheets and Instructions FOXY Probes Probe Construction The standard oxygen probe has its distal tip polished and coated with the o

Strany 121 - FOXY-AL300-000-02-0207 111

6: Hardware Data Sheets and Instructions Sampling Chambers The oxygen probe should be installed in a suitable sampling compartment. If you are not u

Strany 122 - Manager

6: Hardware Data Sheets and Instructions Temperature and the Oxygen Probe Temperature affects the fluorescence decay time, the fluorescence intensit

Strany 123 - FOXY-AL300-000-02-0207 113

6: Hardware Data Sheets and Instructions Use Warnings • Handle with care. Dropping the probe may cause permanent damage. • Gently remove the plastic

Strany 124 - 114 FOXY-AL300-000-02-0207

6: Hardware Data Sheets and Instructions Method Results ETO (at room temperature) Unknown. ETO (at temperatures above 100 °C) Unknown. Hydrogen Pe

Strany 125 - External Triggering

6: Hardware Data Sheets and Instructions FOXY-R Probe FOXY-R Specifications Specification Value Fiber core 1000 µm core silica Fiber cladding Sili

Strany 126 - 116 FOXY-AL300-000-02-0207

6: Hardware Data Sheets and Instructions FOXY-AL300 Aluminum-jacketed Fiber Optic Probe The FOXY-AL300 is a 300-µm aluminum jacketed fiber optic pro

Strany 127 - Chemical Effects on Oxygen

6: Hardware Data Sheets and Instructions Specification Value Maximum applied pressure 300 psi Probe temperature range -80° C to 80° C (reflects ran

Strany 128 - 118 FOXY-AL300-000-02-0207

About This Manual Document Purpose and Intended Audience This document provides you with an installation and configuration instructions to get your

Strany 129 - Overview

6: Hardware Data Sheets and Instructions FOXY-PI600 Specifications Specification Value Fiber core 600 µm silica Fiber coating 25 µm polyimide Outer

Strany 130 - 120 FOXY-AL300-000-02-0207

6: Hardware Data Sheets and Instructions Foxy 21G (left) and 18G (right) Fiber Optic Probe FOXY-18G/21G Specifications Specification Value Fiber c

Strany 131 - FOXY-AL300-000-02-0207 121

6: Hardware Data Sheets and Instructions FOXY-OR125 and FOXY-OR125G O-ring Fiber Optic Probes The FOXY-OR125 is a silica-core, 1000-µm stainless ste

Strany 132 - Model # 4

6: Hardware Data Sheets and Instructions Specification Value Probe temperature range -80° C to 80° C (reflects range of ruthenium complex in the sol

Strany 133 - FOXY-AL300-000-02-0207 123

6: Hardware Data Sheets and Instructions FOXY-T1000 Specifications Specification Value Fiber core 1000 µm silica Fiber cladding Silica Fiber jacket

Strany 134 - 124 FOXY-AL300-000-02-0207

6: Hardware Data Sheets and Instructions FOXY-T1000-RTD Stainless Steel Probe with Light Shield FOXY-T1000-RTD Specifications Specification Value

Strany 135 - FOXY-AL300-000-02-0207 125

6: Hardware Data Sheets and Instructions FOXY-RESP Respiration Probe The FOXY-RESP Respiration Probe is typically used for measuring oxygen tension

Strany 136 - 126 FOXY-AL300-000-02-0207

6: Hardware Data Sheets and Instructions Specification Value Compensation Temperature Probe reconditioning N/A – disposable sensor films Re-calibra

Strany 137 - FOXY-AL300-000-02-0207 127

6: Hardware Data Sheets and Instructions Sample spectra indicative of a compromised probe overcoat FOXY-AF and FOXY-AF-MG Silicone Overcoats The FO

Strany 138 - 128 FOXY-AL300-000-02-0207

6: Hardware Data Sheets and Instructions Sterilizing See the Sterilization table in the Oxygen Sensing Probes – General Information section of this