External Triggering Options Instructions
Overview
Ocean Optics USB2000, USB2000+, USB4000, HR2000, HR2000+, HR4000, Jaz, Maya, QE65000, NIR
Series, NIRQuest Series, S2000, and S1024DW Spectrometers provide several methods of acquiring data
(see table below). In the Normal mode, the spectrometer is “free running.” That is, the spectrometer is
continuously scanning, acquiring, and transferring data to your computer, according to parameters set in
the software. In this mode, however, there is no way to synchronize the scanning, acquisition, and transfer
of data with an external event. However, trigger pulses for synchronizing an external event with the
spectrometer are available.
To synchronize data acquisition with external events, other modes of acquiring data are available. Each
mode involves connecting an external triggering device to the spectrometer and then applying an external
trigger to the spectrometer before the software receives the data. The length of the integration time and
the source for the integration clock depend upon the mode chosen. All other acquisition parameters are set
in the software.
For spectrometers with FPGA firmware versions of 3.0 or higher, and for spectrometers not covered by
this document, see
New External Triggering Options Instructions for Spectrometers with Firmware
Version 3.0 and Above. Also see the HR4000 and USB4000 Shutter Mode Performance in Hardware
Trigger Mode for more information on external triggering with these spectrometers.
Triggering Mode Description Applicable Spectrometer(s)
Normal Spectrometer acquires spectra continuously. All
External
Software
Integration time is set in the software. Software
receives a trigger event and transmits spectra obtained
in the data acquisition cycle in which the trigger
occurred.
All
External
Hardware
Integration time set via chip in spectrometer (USB
Series/Jaz/HR Series/S2000 D Series) or jumpers
(S2000 series). Spectrometer waits for a sharp rise in
voltage on the trigger input pin, and then acquires
spectra until the voltage is removed.
USB2000/USB4000/Jaz/
HR2000/HR2000+/HR4000/
S2000/S1024DW
External
Synchronization
Spectrometer acquires data from an external trigger
event (such as a push button) until the next time the
trigger is activated, at which time the spectrometer
ceases spectral acquisition and begins a new
acquisition. Integration time cannot be set, since the
trigger can fire at random intervals.
S2000/S1024DW
Zobrazit stránku 0
1 2 3 4 5 6 ... 13 14

Shrnutí obsahu

Strany 1 - Overview

200-00000-000-01-201401 1 External Triggering Options Instructions Overview Ocean Optics USB2000, USB2000+, USB4000, H

Strany 2 - 2 200-00000-000-01-201401

External Triggering Options Instructions 10 200-00000-000-01-201401 External Hardware Triggering – Trigger Timing When to Use External Hardware T

Strany 3 - Triggering an External Event

External Triggering Instructions 200-00000-000-01-201401 11 If you enable this function, you will be presented with a file save dialog box with

Strany 4 - Edge Triggering

External Triggering Options Instructions 12 200-00000-000-01-201401 When to Use External Hardware Trigger Mode Use the External Hardware Trigger mo

Strany 5 - Software Support

External Triggering Instructions 200-00000-000-01-201401 13 S1024DW Miniature Fiber Optic Spectrometer Integration Time SAD500

Strany 6 - Palm Spec

External Triggering Options Instructions 14 200-00000-000-01-201401 7. Enable the Automatically save file on trigger box if you wish to save the d

Strany 7 - 200-00000-000-01-201401 7

External Triggering Options Instructions 2 200-00000-000-01-201401 Triggering Mode Description Applicable Spectrometer(s) Quasi-Real Time Acquisi

Strany 8 - S1024DW Spectrometers Only

External Triggering Instructions 200-00000-000-01-201401 3 Trigger Input Pin Information for Ocean Optics Spectrometers Trigger Mode Trigger Inpu

Strany 9

External Triggering Options Instructions 4 200-00000-000-01-201401 Level Triggering and Edge Triggering In External Triggering, there are two modes

Strany 10 - OOIBase32

External Triggering Instructions 200-00000-000-01-201401 5 Integration Times Available with External Triggering The following table illustrates t

Strany 11 - Spectrometers

External Triggering Options Instructions 6 200-00000-000-01-201401 Product Supported Not Supported OOISensors External Software External Synchron

Strany 12

External Triggering Instructions 200-00000-000-01-201401 7 Using the External Software Trigger Mode ► Procedure 1. Connect the line to the appr

Strany 13

External Triggering Options Instructions 8 200-00000-000-01-201401 Once you select an external trigger mode, your computer will appear unresponsive

Strany 14 - 14 200-00000-000-01-201401

External Triggering Instructions 200-00000-000-01-201401 9 Using the External Synchronization Trigger Mode ► Procedure 1. Supply a line from Pi

Komentáře k této Příručce

Žádné komentáře